RSS Feed
News
Jun
23
Adding support for M9256 Flow Transducer lots 101 onward.
Posted by Esko Luostarinen on 23 June 2015 11:11 AM

Dear Medikro Spirometry Software user

Description

Medikro Spirometry Software versions 3.0 to 3.0.3 do not support using M9256 Disposable Flow Transducer lot numbers 101 onwards. To use the flow transducers mentioned, you have to run a patch file, which adds the support to Medikro Spirometry Software. If you have Medikro Spirometry Software v3.1, M9256 lots 101 onwards are already supported and no actions are needed. Naturally in this case you can also purchase a software update, and benefit from new features like the GLI reference set, and the brand new Medikro Calibration Free Technology.

Applies to

Medikro Spirometry Software versions 3.0, 3.0.1, 3.0.2 and 3.0.3.

Instructions

  1. Download a patch file M9888_Patch_for_MSS_(UserNote_M5236).exe from Medikro eSupport: Knowledgebase->Downloads->Patch files. The patch is a self-extracting program that adds the support automatically. Patch can be used with Medikro Spirometry Software versions 3.0 to 3.0.3 only. 
  2. Run the patch by double-clicking the M9888_Patch_for_MSS_(UserNote_M5236).exe.
  3. If the operating system asks: “Do you want to allow the following program to make changed to this computer? Patch for Medikro Spirometry Software (User Note M5236)” answer “Yes”.
  4. After running the patch, program will show a message of successful operation. Click “OK”
  5. You have to restart the Medikro Spirometry Software.

In case you have any questions regarding this patch, or if you require any assistance, please submit a Ticket with Medikro eSupport.


Comments (0)